ES-3597 Rev.5

KWLD

Plug-in Signal Conditioners K-UNIT

CURRENT LOOP SUPPLY

(for level transmitter use)
Functions & Features
• Powering a 4 – 20 mA DC current loop
• Shortcircuit protection
• Applicable to smart transmitters
• Usable as 4 – 20 mA isolator
• Wide span adjustment range
• High-density mounting

MODEL: KWLD–A6–[1][2]

ORDERING INFORMATION

• Code number: KWLD-A6-[1][2]
Specify a code from below for each of [1] and [2].
  (e.g. KWLD-A6-B/Q)
• Specify the specification for option code /Q
  (e.g. /C01/S01)

INPUT

Current
4 – 20 mA DC (Input resistance 250 Ω)

OUTPUT 1

Current
A: 4 – 20 mA DC (Load resistance 600 Ω max.)

OUTPUT 2

Voltage
6: 1 – 5 V DC (Load resistance 5000 Ω min.)

[1] POWER INPUT

AC Power
B: 100 V AC
C: 110 V AC
D: 115 V AC
F: 120 V AC
G: 200 V AC
H: 220 V AC
J: 240 V AC
DC Power
S: 12 V DC
R: 24 V DC

[2] OPTIONS

blank: none
/Q: With options (specify the specification)

SPECIFICATIONS OF OPTION: Q (multiple selections)

COATING (For the detail, refer to our web site.)
/C01: Silicone coating
/C02: Polyurethane coating
/C03: Rubber coating
TERMINAL SCREW MATERIAL
/S01: Stainless steel

GENERAL SPECIFICATIONS

Construction: Plug-in
Connection: M3.5 screw terminals
Screw terminal: Chromated steel (standard) or stainless steel
Housing material: Flame-resistant resin (black)
Isolation: Input to output 1 to output 2 to power
Overrange output: Approx. -10 to +110 %
Zero adjustment: ±10 % of the input 4 – 20 mA (front)
Output 2 zero adjustment: ±2 % of the calibrated range (front)
Span adjustment: 10 – 100 % of the input 4 – 20 mA (front)
Output 2 span adjustment: ±2 % of the calibrated range (front)
Coarse span adjustment: Rotary switch (front)
Loop test switch: Supply output is removed with OFF.

SUPPLY OUTPUT

(Across the terminals 4 – 5)
Output voltage: 24 – 28 V DC with no load
18 V DC min. at 20 mA
Current rating: ≤ 22 mA DC
Shortcircuit Protection
Current limited
: 30 mA max.
Protected time duration: No limit

INPUT SPECIFICATIONS

DC Current: Input resistor incorporated

INSTALLATION

Power input
AC: Operational voltage range: rating ±10 %,
50/60 ±2 Hz, approx. 3.5 VA
DC: Operational voltage range: rating ±10 %
ripple 10 %p-p max., approx. 3 W (125 mA at 24 V)
Operating temperature: -5 to +55°C (23 to 131°F)
Operating humidity: 30 to 90 %RH (non-condensing)
Mounting: Surface or DIN rail
Weight: 400 g (0.88 lb)

PERFORMANCE in percentage of span

Accuracy: ±0.1 % (Coarse span adjustment SW set to ‘0’)
Temp. coefficient: ±0.02 %/°C (±0.01 %/°F) (Coarse span adjustment SW set to ‘0’)
Response time: ≤ 0.2 sec. (0 – 90 %)
Line voltage effect: ±0.1 % over voltage range
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength:
2000 V AC @1 minute (input to output 1 or output 2 to power to ground)
1000 V AC @1 minute (output 1 to output 2)

EXTERNAL VIEW

EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm [inch]

SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM