ES-0173 Rev.0

LCCT-2-1

STRAIN GAUGE LOAD CELL

(Tension and compression type)
Functions & Features
• Rated capacity 2 N
• Very small type
• Easy to install with female screws on both ends

MODEL: LCCT-2-1

ORDERING INFORMATION

• Code number: LCCT-2-1

GENERAL SPECIFICATIONS

Cable type: 1.6 dia., 4-core shielded cable,
split wire at the end
Cable length: 1 m
Body material: beryllium copper
Rated flexure: 0.02 mm or less
Natural frequency: Approx. 4.9 kHz

INPUT SPECIFICATIONS

Rated capacity (R.C.): 2 N (204 gf)
Safe overload: 200 % R.C.
Recommended excitation voltage: 5 V
Maximum excitation voltage: 10 V
Input terminal resistance: Approx. 1000 Ω

OUTPUT SPECIFICATIONS

Rated output (R.O.): 0.5 to 1.5 mV/V
Output terminal resistance: Approx. 1000 Ω

INSTALLATION

Compensated temperature range: -10 to +50 ℃ (14 to 
122 °F)
Safe temperature range: -10 to +60 ℃ (14 to 140 °F)
Weight: 3 g (0.11 oz)

PERFORMANCE in percentage of span

Zero balance: ±50 % R.O.
Non-linearity: ±0.2 % R.O.
Hysteresis: ±0.2 % R.O.
Repeatability: ±0.1 % R.O.
Temperature effect on zero: ±0.1 % R.O./℃
Temperature effect on output: ±0.1 % LOAD/℃
Insulation resistance: 1000 MΩ or more (B/E 50V DC)

EXTERNAL DIMENSIONS unit: mm [inch]